vol.38 número3An investigation into specialist practice nurses' knowledge of cardiopulmonary resuscitation guidelines in a tertiary hospital in Gauteng Province, South AfricaThe incidence and outcomes of patients with acute kidney injury in a multidisciplinary intensive care unit in Durban, South Africa índice de autoresíndice de materiabúsqueda de artículos
Home Pagelista alfabética de revistas  

Servicios Personalizados

Articulo

Indicadores

Links relacionados

  • En proceso de indezaciónCitado por Google
  • En proceso de indezaciónSimilares en Google

Compartir


Southern African Journal of Critical Care (Online)

versión On-line ISSN 2078-676X
versión impresa ISSN 1562-8264

Resumen

WILSON, K; FOURTOUNAS, M  y  ANAMOURLIS, C. A comparison of the warming capabilities of two Baragwanath rewarming appliances with the Hotline fluid warming device. South. Afr. j. crit. care (Online) [online]. 2022, vol.38, n.3, pp.97-104. ISSN 2078-676X.  http://dx.doi.org/10.7196/SAJCC.2022.v38i3.549.

BACKGROUND. Accidental intraoperative hypothermia is a common and avoidable adverse event of the perioperative period and is associated with detrimental effects on multiple organ systems and postoperative patient outcomes. In a resource-limited environment, prevention of intraoperative hypothermia is often challenging. Resourceful clinicians overcome these challenges through creative devices and frugal innovations. OBJECTIVE. To investigate the thermal performance of two Baragwanath Rewarming Appliances (BaRA) against that of the Hotline device to describe an optimal setup for these devices. METHODS. This was a quasi-experimental laboratory study that measured the thermal performance of two BaRA devices and the Hotline device under a number of scenarios. Independent variables including fluid type, flow rate, warming temperature and warming transit distance were sequentially altered and temperatures measured along the fluid stream. Change in temperature (ΔT) was calculated as the difference between entry and exit temperature for each combination of variables for each warming device. RESULTS. A total of 219 experiments were performed. At a temperature of 43.0°C and a transit distance of 200 cm, the BaRA A configuration either matched or exceeded the ΔT of the Hotline over all fluid type and flowrate combinations. The BaRA B configuration does not provide comparable thermal performance to the Hotline. Measured flowrates were noticeably slower than manufacturer-quoted values for all intravenous (IV) cannulae used. CONCLUSION. A warm-water bath at 43.0°C with 200 cm of submerged IV tubing provides thermal performance comparable to the Hotline device, with all fluid type and flowrate combinations.

Palabras clave : perioperative hypothermia; frugal innovation; warming devices.

        · texto en Inglés     · Inglés ( pdf )

 