Water SA vol.46 n.3 Pretoria Jul. 2020
http://dx.doi.org/10.17159/wsa/2020.v46.i3.8661
RESEARCH PAPERS
Development of a paper-based microfluidic device for the quantification of ammonia in industrial wastewater
Nonhlazeko Loveday NxumaloI; Lawrence Mzukisi MadikizelaI; Hendrik Gert KrugerII; Stanley Chibuzor OnwubuIII; and Phumlane Selby MdluliI
IDepartment of Chemistry, Durban University of Technology, PO Box 1334, Durban, 4001, South Africa
IICatalysis and Peptide Research Unit, School of Health Sciences, Westville campus, University of KwaZulu-Natal, Durban, 4000, South Africa
IIIDental Sciences, Durban University of Technology, PO Box 1334, Durban, 4001, South Africa
ABSTRACT
Ammonia is a toxic pollutant increasingly found in urban and industrial wastewater and unprotected surface water. Industry discharges and fertilizer run-off release ammonia into sewers and streams, overloading wastewater treatment plants and causing fish deaths in surface water such as rivers, sea and lakes. The purpose of this study was to develop and evaluate the effectiveness of the microfluidic paper-based device (µPAD) for the quantification of ammonia in wastewater. The µPAD fabricated had an oval-shaped pattern which was designed using CorelDraw software. The hydrophilic zones were created by printing a chromatographic paper with a Xerox wax printer (Xerox colorqube 8570). The modified version of the colorimetric method using Nessler reagent was combined with microfluidic technologies to create a low-cost monitoring system for detection of ammonia in wastewater. The method allows for ammonia determination in the range of 0-5 ppm (mg/L) with a limit of detection of 3.34 ppm. This study indicated that a µPAD was successfully used to quantify the concentration of ammonia in wastewater.
Keywords: ammonia, microfluidic device, quantification
INTRODUCTION
Ammonia (NH3) is a chemical with widespread application/production in different industries. It is the most commonly occurring nitrogenous pollutant in wastewater (Lee et al., 2009). Ammonia can be present as either gaseous NH3 or ammonium ion (NH4+) in aqueous solutions, depending on solution pH. The presence of ammonia in wastewater is regarded as a major concern due to its toxicity. Toxicity of ammonia can be problematic to fish and other aquatic animals, resulting in the depletion of dissolved oxygen (Colt and Huguenin, 2002) as well as eutrophication (Jorgensen, 2002). It is already reported that a concentration as low as 0.2 ppm (mg/L) can be fatal to aquatic organisms (Wiesmann, 1994).
The guideline levels for ammonia concentrations released into receiving water bodies vary in different parts of the world. The United States Environmental Protection Agency (USEPA) and the American Committee on Water Quality Criteria recommend a value of below 0.02 ppm (mg/L) nitrogen-ammonia (Jorgensen, 2002). Therefore, regulatory agencies require industries to treat ammonia in their effluents to meet these stringent requirements. However, some existing wastewater treatment plants (WWTP) are unable to comply with these regulations as their service load increases. Industries are also experiencing increased regulations on their wastewater and are being cut-off from sewer discharge.
Currently, monitoring ammonia in wastewater is primarily based on manual sampling followed by analysis using standard laboratory methods. This results in the infrequent monitoring of wastewater at a much lower number of locations than is desirable, as it is time-consuming, expensive, non-scalable and requires skilled personnel. Therefore, it is of great importance to seek an efficient, practical, time- as well as cost-effective, and sensitive method for the detection of ammonia in wastewater.
Numerous methods for the identification and quantification of ammonia such as potentiometric (Jayawardane et al., 2015), colorimetric (Amirjani and Fatmehsari, 2018) and titrimetric (Thompson and Morrison, 1951) measurements have been reported in literature. However, these analytical methods each have their own restrictions, and different levels of investment in equipment, operating costs, secondary pollution, and other deficiencies (Guo et al., 2008; Islam and Patel, 2010; Luo et al., 2015; Markou et al., 2014; Yang and Cheng, 2007).
Consequently, there has been great interest in developing a portable, sensitive and optical ammonia sensor which is operable at ambient conditions and capable of detecting ammonia at parts-per-million levels (Srivastava et al., 2012). It has been suggested that a low-cost microfluidic paper-based analytical device offers an opportunity to tackle this need, thereby increasing the frequency and geographic coverage of environmental monitoring while also reducing analytical costs and complexity of measurement (Meredith et al., 2016).
Microfluidic paper-based analytical devices (µPADs) have been developed using various patterning technologies to reproducibly create low-cost lab-on-chip types of analytical tools (Yamada et al., 2015). Consequently, these devices are useful for field-deployable measurements in the areas of healthcare diagnostics, environmental monitoring and forensic analysis. These devices are portable, cost-effective and easily reproducible (Gong and Sinton, 2017; Martinez et al., 2010). One of the significant characteristics of µPADs is that they are extremely flexible. Thus, they can be adjusted to any application, particularly in low-resource settings.
The most widely used analytical method performed with paper devices is the simple yet effective colourimetric assay. Colorimetric detection is a well-established, fundamental method that has been the most widely adopted sensing mechanism for microfluidics paper-based analytical devices (Nie et al., 2012). This paper describes the use of the developed microfluidic paper-based analytical device (µPAD) combined with colorimetric detection chemistries for the quantification and detection of ammonia in wastewater. In particular, the most commonly used spectrophotometric color reaction in ammonia monitoring is that of Nessler reagent (K2HgI4). This method is the preferred method and has the advantages of easy operation and high sensitivity. The colorimetric detection of ammonia includes the use of a Nessler reagent and salicylate method. In the Nessler reagent method, mercury (II) chloride and potassium iodide react with ammonia to form a yellow-brown coloured compound (Koch and McMeekin, 1924).
The XYZ tristimulus values unfortunately have constrained use as colour descriptions since they correspond poorly with visual traits. While Y relates to value (lightness), X and Z do not correlate to hue and chroma. It is therefore not particularly perceptually uniform (Baldevbhai and Anand 2012). Thus, when the 1931 CIE (Fairchild, 1993) standard observer was established, the commission endorsed using the chromaticity coordinates L∗a∗b∗/Yxy. Since the smartphone camera acquires images in the RGB colour space, the smartphone image information of RGB values was processed to obtain L∗, a∗ and b∗ values using an empirical colour space conversion model (Hunter Lab) (León et al., 2006; Wyszecki, 1968). The CIELAB colour space has a uniform colour distribution, and is very close to human perception of colour. The colour is represented by a point; this therefore allows high precision. More so, L∗a∗b∗ values do not elucidate absolute colours except if the white reference is also specified. Frequently, the white reference is assumed to follow a standard and is not directly specified (Consortium, 2004).
Colorimetry
The addition of Nessler reagent to the solution containing ammonia leads to a change in the appearance of the solution, i.e., the colourless initial solution converts to intense yellow. The formation of NH2Hg2I3 compound is responsible for this colour change which provides real-time and observable monitoring of the system. It was then possible to automatically capture images of the observed colour. After capturing, the required images were measured using computer software, i.e., Adobe photoshop and ImageJ. The mean gray value (intensity) and the RGB (red, green, blue) values were extracted using ImageJ software.
MATERIALS AND METHODS
Chemicals and apparatus
Mercury (II) chloride (99 %), potassium sodium tartrate and sodium nitroferricyanide (99%) were purchased from Sigma-Aldrich (Steinheim, Germany). Potassium iodide, sodium hydroxide (99%), sodium salicylate and ammonium chloride were purchased from Merck (Darmstadt, Germany). A Cary 50 Conc UV-visible spectrophotometer purchased from SMM Instruments PTY (LTD) (Cape Town). A Samsung Galaxy J1 smartphone was purchased from Game (Durban, South Africa).
Preparation of reagents
Nessler's reagent
Nessler reagent was prepared by dissolving 4.4 g of mercury (II) chloride in 40 mL distilled water. Complete dissolution was achieved by heating the solution on the hot plate. In a separate flask, 11.0 g of potassium iodide was dissolved in the minimum possible amount of water (approximately 2 mL). This saturated solution of potassium iodide (KI) was then added drop-wise into the prepared solution of mercury (II) chloride. This resulted in abundant precipitation of orange mercury (II) Iodide (HgI2). A 100 mL solution of 5 M sodium hydroxide was prepared by dissolving 20 g of sodium hydroxide in sufficient water. Thereafter, 40 mL of the prepared sodium hydroxide solution was transferred into the complex solution in a 250 mL beaker.
Ammonia standard solutions
A stock solution containing 100 ppm (mg/L) of ammonia was prepared by dissolving 0.3818 g of ammonium chloride in distilled water and diluting to 1 000 mL with deionised water. Standard solutions in the concentration range of 0-5 ppm (mg/L) of ammonia were prepared by suitable dilutions of the prepared stock solution.
Experimental
Fabrication of the µPAD
The patterning of the microfluidic paper-based analytical device was designed using CorelDraw X7 software, followed by wax printing on chromatographic paper using a wax printer (Xerox Colorqube 8570). In general, identical patterns of wax were printed in alignment with the Whatman No.1 chromatography paper substrate which was cut to A4 size before use. The sheets of Whatman No.1 chromatography paper were directly fed into the printer. The print head dispenses ink (melted wax) on the surface of the paper, where it cools and solidifies instantly without further spreading. The ink is made of a mixture of hydrophobic compounds including carbamates, hydrocarbons and dyes that melt at around 120°C. The μPAD were produced by the wax pattern manufacturing process.
The fabricated μPAD was then used for the colorimetric detection of ammonia using Nessler's reagent method (the mixture of K2HgI4 and NaOH or KOH). The analysis of ammonia was achieved by dropping 10 µL Nessler reagent and 5 µL of potassium sodium tatrate into a reagent zone of the µPAD containing 20 µL of ammonia standard solution of different concentrations ranging from 0 to 5 ppm (mg/L).
Smartphone imaging
Images of the μPAD were taken with a smartphone under fluorescent lighting conditions (laboratory lights). The smartphone camera light source was set on its continuous operation mode with flash off, a resolution of 3.9 M pixels (white balance, white fluorescent) and ISO camera speed of 100. The camera was focused on the μPADs placed on a white bench surface; the images were then captured after 1 min of the observed colour change. The distance between the μPAD and the camera was fixed to 20 cm and the captured images were thereafter quantitatively analysed using ImageJ software.
Analytical procedure
The fabricated µPAD was evaluated for determination of ammonia in wastewater collected from Acacia effluent treatment plant (ETP) dam over 10 consecutive days. Acacia operation services is situated at Umbogintwini Industrial Complex (UIC) and is the owner of the complex and the only industry licensed to discharge wastewater to the Kingsway Sea. Aqueous samples were collected as a composite sample. A mixture of samples was collected at 3 intervals: morning, evening and midnight. All samples for laboratory analysis were placed into thoroughly cleaned 1 L plastic bottles and tightly closed. Each bottle was rinsed with the appropriate amount of sample before final sample collection. These samples were placed in a cooler box, protected from direct sunlight, and taken to the laboratory for analysis. The µPAD results for samples that did not contain suspended particulate matter and coloured compounds were compared with those obtained by the standard spectrophotometric salicylate method.
The analytical procedure was conducted at room temperature and involved the deposition of 10 µL of sample or standard solutions into the individual sample detection zones of the µPAD, followed by the addition of 5 µL of Nessler reagent. Addition of the Nessler reagent resulted in the colour change. The colour change was captured with a smartphone, and thereafter transferred to a computer for colour measurement. ImageJ software (National Institute of Health USA, http://imagej.nih.gov./ij) was used to measure the intensity of the colour formation in the detection zones. The RBG (red, green, blue) colour intensity profile plot was obtained for a chord passing through the centre of each detection zone. The highest sensitivity was obtained for green and red colours. This method was then validated by measuring the optical characteristics using a UV-visible spectrophotometer.
RESULTS AND DISCUSSION
Optimization results
The results obtained indicated that the intensity of the colour was directly proportional to the concentration of ammonia present in aqueous solution. The developed colours corresponding to various concentrations of ammonia are represented in Fig. 1. It can be observed that low ammonia concentrations resulted in the formation of a light-yellow colour on the µPAD. The yellow colour became more intense at higher ammonia concentrations.
The experimental data obtained proved that the Nessler reagent was soluble in water and provided a clear colour change from colourless to yellow. The effect of the concentration of the Nessler reagent on the intensity value was studied in the range from 0-5 ppm. The intensity value increased rapidly with increasing concentration of ammonia.
A calibration curve (Fig. 2) for the colorimetric detection of ammonia with the Nessler reagent µPAD was successfully obtained by a smartphone thus showing that colour intensity measurements of µPADs could be conducted in the field. The analytical performance of the Nessler reagent was evaluated and the corresponding data are summarized in Table 1. Deionsed water was used as the blank. All colour intensity measurements were quadruplicated. The calibration curves of the Nessler reagent µPADs for their corresponding linear ranges (0-5 ppm) are described in Table 1. The limit of detection (n ≥ 10) of the method was determined by the linear regression method of Miller and Miller.
Colour measurements
In order to assess the possibility to use the device outside a laboratory setting, digital images of the paper devices were taken using a smartphone. The use of a smartphone would allow the user to be less reliant on expensive equipment and an established laboratory to complete the analysis (Moonrungsee et al., 2015). For example, to be able to use the paper device in the field the images of the paper devices could be sent via email to a central laboratory and analysed or the ImageJ software used to analyse the data from the paper device could be turned into a smartphone-based application (app) and analysed immediately using the mobile phone (Lopez-Ruiz et al., 2014; Shen et al., 2012). In this study the colour accuracy of the images captured for the µPADs in the colorimetric detection of ammonia using Nessler reagent was determined by measuring the RGB (red, green and blue) values and using CIELab (Commission Internationale de l'Eclairage) colour space determination.
RGB colour measurement
The RGB colour model is based on the science of how the human eye peceives light and translates it into signals (Alala et al., 2014). The colorimetric image of ammonia detected was analysed for RGB values using ImageJ software. The obtained RGB colour histograms showing the pixels and the RGB values of each wastewater sample, performed in duplicate, are depicted in Fig 3. From a close look at the RGB pixels, it is seen that, for the more intense yellow-coloured samples the R and G values were shifting closer to the region of 225, and the B value closer to 0.
The RGB values and grayscale showed no monotonic change in response to the concentration value of 0-5 ppm and narrow detectable concentration ranges with RGB values. This is due to the lack of colour information because the respective RGB values show only the information of redness, greenness or blueness of the resulting colours.
CIELAB colour measurement
From the RGB colour modeling, the RGB values were converted to L∗a∗b∗ using Hunter Lab in order to obtained the CIE L∗a∗b∗/Yxy colour space, a colour space (L∗a∗b∗) in which L represents lightness, a∗ represents the chromatic coordinates for red-green (+a∗ = red direction, −a∗ = green direction) and b∗ represents the chromaticity coordinate for yellow-blue (+b∗ = yellow direction, −b∗ = blue direction). This colour space includes all perceivable colours it has a gamut that is greater than human vision. The chromatic diagrams for the CIEL Yxy for the detected colour of each standard are shown in Fig. 4.
The chromaticity diagrams in Fig. 4 for quadruplicate analysis of each standard (from 0 to 5 ppm) showed that the colours were becoming more intensely yellow with the increase in concentration. The image analysis performed with the CIE L∗a∗b∗ colour coordinates (L∗, a∗, and b∗) (Fig. 4) indicated a trend in the data, accompanying an increase in concentration. It also shows dependencies of the concentration on each coordinate (L∗, a∗, and b∗). There is a shift in the colour space coordinates of the μPAD images due to different ambient light conditions.
The CIELab colour difference between the predicted and measured shade tab colours were then used as indices of accuracy for the camera. Colour difference (∆E) in Eq. 1 is defined by the following equation:
where ∆L∗, ∆a∗ and ∆b∗ are the respective differences between the measured and predicted CIELAB values of the shade tabs. The colour difference for each standard was compared to that of a blank. The results for the calculated ∆E for each standard are shown in Fig. 4. If ∆E is between the values of 0.5 and 1.0, the difference is perceivable for experienced observers only and if ∆E ranges between 1 and 2, it is said to be the minimal colour difference. A clear colour difference can be observed when ∆E is between 3 and 5 (Mokrzycki and Tatol, 2011). The calculated ∆E for each standard in Fig. 5 shows that the colour difference for 0.1 ppm is normally undetectable, while for 0.5 ppm there is a very small difference, only obvious to a trained eye; 1 ppm shows an obvious difference and 3-5 ppm shows a very obvious difference.
The ∆E plot for each standard in Fig. 5 showed that as concentration increased, the colours of the standards were significantly different at each concentration. Thus, it was observed that there was an increase in colour with the increase in concentration.
Analysis of wastewater samples
The image from the colorimetric detection of ammonia in ETP dam samples is shown in Fig. 6. The results obtained in the evaluation of the Nessler reagent µPADs for the determination of ammonia in wastewater samples are summarised in Table 2. The concentration of ammonia varies for each sample as shown in Fig. 6.
Traces of ammonia were detected in the ETP dam samples in ppm concentrations. The concentrations obtained from these samples were within the optimization range for the ammonia standards (0-5 ppm). This finding is consistent with Luo et al. (2015) who reported a detection range of 0.1-10 ppm in wastewater samples.
Varying ammonia concentrations were detected from each sample. The Nessler method, with the change in colour intensity from pale yellow to intense yellow (Fig. 6) suggested the presence of ammonia. This finding was further supported by the previous work of Jeong and co-workers that showed that the presence of ammonia results in a change in colour from pale yellow to intense yellow (Jeong et al., 2013). The concentration of ammonia in samples taken over 10 days was found to be dissimilar. The traces of ammonia detected in UIC wastewater samples were in ppm concentration units (Table 2). For Days 1, 2, 4, 5, 6, and 7 there was a visible colour change observed. The change in colour observed from these samples indicated the presence of ammonia with each method. Martinez-Carballo and co-workers suggested that the detected ammonia concentrations may be due to the matrix effects being higher in the influents than in the effluents of WWTPs, which led to higher matrix suppression and higher quantification limits in influents (Martínez-Carballo et al., 2007). In particular, naked eye observation of colour change for Days 3, 8, 9, and 10, showed that there was a slight colour change, which was not visible enough. Therefore, the ammonia concentration detected from these samples was low. The colorimetric analysis of the Nessler reagent indicated that the resulting ammonia concentrations obtained from the UIC ETP dam wastewater samples was at the level of that of domestic wastewater, which indicates that it is permissible for it to be discharged directly to a sewer, as far as the nitrogen concentration is concerned.
Results presented in Fig. 7 show the colorimetric quantification of ammonia in a sample from the UIC ETP dam using the Nessler reagent method. It was observed that although the concentration of ammonia detected from Days 1-10 of the ETP samplingvaries, a gradual decrease was also noticed. An increase in ammonia level indicated a change in toxicity or pH. The experimental data gathered confirmed that μPAD was highly successful in the quantification of ammonia.
Interferences
To evaluate the combined effect of inorganic ions present in wastewater (Ca2+, Mg2+, Fe2+, Ni2+, and Cu2+) on the determination of ammonia, doubly concentrated 'very hard water', according to a standard recipe, was tested (Verdouw et al., 1978). The effect of potential interferences on the detection of ammonia in wastewater was studied by spiking each interference ion into the ammonia sample. The observed colour change image is depicted in Fig. 8.
It was observed that for Ca2+ there was no intense yellow colour; instead, a dark brown colour with gelatinous black precipitate was observed. For Ni2+, Cu2+, Ca2+ and Mg2+ an intense yellow colour was observed. The quantitative recoveries of ammonia presented in Table 3 were in the range of 96-106%. These recoveries indicated that the proposed µPAD can be used for the determination of ammonia in wastewater samples. Accurate analysis of ammonia content using the Nessler method cannot be possible when Ca2+ is present in wastewater samples. Ca2+ in wastewater is known to precipitate and forms a complex with salts in the Nessler reagent. The modified Nessler method could not accurately quantify ammonia in water samples with Ca2+. Therefore, a stabilizer solution of potassium sodium tartrate should be added to halt the development of cloudiness of the Nessler reagent which may occur in some wastewaters because of calcium interference.
Applicability of the colorimetric method
The accuracy of the proposed method in this paper was comparable with that of previously published techniques. It is worthwhile to note that the linear ranges presented in Table 4 are wider than those measured for the proposed method. Also, this method holds the advantages of simplicity and rapidity as well as the fact that microfluidic paper-based detection of ammonia is a low-cost analytical method which can detect even the smallest concentration of the analyte.
CONCLUSIONS
Although µPADs were originally developed as biomedical assays for point-of-care diagnostics (Chong et al., 2013), their use has expanded into environmental research. Relative to the progress achieved in biomedical devices, environmental monitoring with µPADs is still in its infancy. The microfluidic paper-based analytical device for the direct determination of ammonia in wastewater has been successfully developed. This paper has reported the optimisation parameters of a simplified Nessler reagent to facilitate the integration of the device into an analysis system, resulting in a simple procedure for the real-time measurement of ammonia in wastewater, with a lower limit of detection achieved of 3.37 ppm.
The feasibility of detecting coloured solutions based on the image colour analysis method was studied. The results showed that the µPAD permitted the method's simple design and easy operation, and provided a simple method to detect the concentration of ammonia in wastewater through the application of computer image processing technology and colour analysis technology. The RGB and the CIELAB colour space were used for computational colour analysis, and the colour difference (∆E) calculation proved that for each standard the colour is different.
Compared with traditional methods, the proposed method has the advantages of high automation, low cost, simple instrument design, and a user-friendly interface. This is suitable for fast water quality detection in the field, and has wide application prospects for measurement and computer science.
REFERENCES
ALALA B, MWANGI W and OKEYO G (2014) Image representation using RGB color space. Int. J. Innov. Res. Dev. 3 (8) pp. [ Links ]
AMIRJANI A and FATMEHSARI DH (2018) Colorimetric detection of ammonia using smartphones based on localized surface plasmon resonance of silver nanoparticles. Talanta 176 242-246. https://doi.org/10.1016/j.talanta.2017.08.022 [ Links ]
BALDEVBHAI PJ and ANAND R (2012) Color image segmentation for medical images using L∗ a∗ b∗ color space. IOSR J. Electron. Commun. Eng. 1 (2) 24-45. https://doi.org/10.9790/2834-0122445 [ Links ]
BOO H-K and MA T (1976) A simple method for determining ammonia in water at the ppm level. Microchim. Acta 66 (5-6) 515-523. https://doi.org/10.1007/BF01220109 [ Links ]
CHONG H, KOO Y, COLLINS B, GOMEZ F, YUN Y and SANKAR J (2013) Paper-based microfluidic point-of-care diagnostic devices for monitoring drug metabolism. J. Nanomed. Biotherm. Discovery 3 e122. https://doi.org/10.4172/2155-983X.1000e122 [ Links ]
COLT J and HUGUENIN JE (2002) Design and Operating Guide for Aquaculture Seawater Systems. Elsevier, London. [ Links ]
CONSORTIUM IC (2004) Image technology colour management - architecture, profile format, and data structure. Specification ICC. 1: 2004-10 (Profile version 4.2. 0.0). [ Links ]
DUONG HD and RHEE JI (2014) A ratiometric fluorescence sensor for the detection of ammonia in water. Sens. Actuators B: Chem. 190 768-774. https://doi.org/10.1016/j.snb.2013.09.071 [ Links ]
FAIRCHILD MD (1993) The CIE 1931 standard colorimetric observer: Mandatory retirement at age 65? Color Res. Appl. 18 (2) 129-134. https://doi.org/10.1002/col.5080180210 [ Links ]
GONG MM and SINTON D (2017) Turning the page: advancing paper-based microfluidics for broad diagnostic application. Chem. Rev. 117 (12) 8447-8480. https://doi.org/10.1021/acs.chemrev.7b00024 [ Links ]
GUO Z, ZHENG Z, GU C and ZHENG Y (2008) Gamma irradiation-induced removal of low-concentration nitrite in aqueous solution. Radiat. Phys. Chem. 77 (6) 702-707. https://doi.org/10.1016/j.radphyschem.2007.10.002 [ Links ]
ISLAM M and PATEL R (2010) Synthesis and physicochemical characterization of Zn/Al chloride layered double hydroxide and evaluation of its nitrate removal efficiency. Desalination 256 (1) 120-128. https://doi.org/10.1016/j.desal.2010.02.003 [ Links ]
JAYAWARDANE BM, MCKELVIE ID and KOLEV SD (2015) Development of a gas-diffusion microfluidic paper-based analytical device (μPAD) for the determination of ammonia in wastewater samples. Anal. Chem. 87 (9) 4621-4626. https://doi.org/10.1021/acs.analchem.5b00125 [ Links ]
JEONG H, PARK J and KIM H (2013) Determination of NH4+ in environmental water with interfering substances using the modified Nessler method. J. Chem. 2013 9. https://doi.org/10.1155/2013/359217 [ Links ]
JORGENSEN TC (2002) Removal of ammonia from wastewater by ion exchange in the presence of organic compounds. Thesis, University of Canterbury, New-Zealand. [ Links ]
KOCH FC and MCMEEKIN TL (1924) A new direct nesslerization micro-kjeldahl method and a modification of the Nessler-Folin reagent for ammonia. J. Am. Chem. Soc. 46 (9) 2066-2069. https://doi.org/10.1021/ja01674a013 [ Links ]
LEE CG, FLETCHER TD and SUN G (2009) Nitrogen removal in constructed wetland systems. Eng. Life Sci. 9 (1) 11-22. https://doi.org/10.1002/elsc.200800049 [ Links ]
LEÓN K, MERY D, PEDRESCHI F and LEÓN J (2006) Color measurement in L∗a∗b∗ units from RGB digital images. Food Res. Int. 39 (10) 1084-1091. https://doi.org/10.1016/j.foodres.2006.03.006 [ Links ]
LOPEZ-RUIZ N, CURTO VF, ERENAS MM, BENITO-LOPEZ F, DIAMOND D, PALMA AJ and CAPITAN-VALLVEY LF (2014) Smartphone-based simultaneous pH and nitrite colorimetric determination for paper microfluidic devices. Anal. Chem. 86 (19) 9554-9562. https://doi.org/10.1021/ac5019205 [ Links ]
LUO X, YAN Q, WANG C, LUO C, ZHOU N and JIAN C (2015) Treatment of ammonia nitrogen wastewater in low concentration by two-stage ozonization. Int. J. Environ. Res. Public Health 12 (9) 11975-11987. https://doi.org/10.3390/ijerph120911975 [ Links ]
MARKOU G, VANDAMME D and MUYLAERT K (2014) Using natural zeolite for ammonia sorption from wastewater and as nitrogen releaser for the cultivation of Arthrospira platensis. Bioresour. Technol. 155 373-378. https://doi.org/10.1016/j.biortech.2013.12.122 [ Links ]
MARTÍNEZ-CARBALLO E, SITKA A, GONZÁLEZ-BARREIRO C, KREUZINGER N, FÜRHACKER M, SCHARF S and GANS O (2007) Determination of selected quaternary ammonium compounds by liquid chromatography with mass spectrometry. Part I. Application to surface, waste and indirect discharge water samples in Austria. Environ. Pollut. 145 (2) 489-496. https://doi.org/10.1016/j.envpol.2006.04.033 [ Links ]
MARTINEZ AW, PHILLIPS ST, WHITESIDES GM and CARRILHO E (2010) Diagnostics for the developing world: microfluidic paper-based analytical devices. Anal. Chem. 82 (1) 3-10. https://doi.org/10.1021/ac9013989 [ Links ]
MEREDITH NA, QUINN C, CATE DM, REILLY TH, VOLCKENS J and HENRY CS (2016) Paper-based analytical devices for environmental analysis. Analyst 141 (6) 1874-1887. https://doi.org/10.1039/C5AN02572A [ Links ]
MOKRZYCKI W and TATOL M (2011) Colour difference∆ E-A survey. Mach. Graphics Vision 20 (4) 383-411. [ Links ]
MOONRUNGSEE N, PENCHAREE S and JAKMUNEE J (2015) Colorimetric analyzer based on mobile phone camera for determination of available phosphorus in soil. Talanta 136 204-209. https://doi.org/10.1016/j.talanta.2015.01.024 [ Links ]
NIE J, ZHANG Y, LIN L, ZHOU C, LI S, ZHANG L and LI J (2012) Low-cost fabrication of paper-based microfluidic devices by one-step plotting. Anal. Chem. 84 (15) 6331-6335. https://doi.org/10.1021/ac203496c [ Links ]
O'CONNOR ŠRAJ L, ALMEIDA MIGS, MCKELVIE ID and KOLEV SD (2017) Determination of trace levels of ammonia in marine waters using a simple environmentally-friendly ammonia (SEA) analyser. Mar. Chem. 194 133-145. https://doi.org/10.1016/j.marchem.2017.06.008 [ Links ]
SHEN L, HAGEN JA and PAPAUTSKY I (2012) Point-of-care colorimetric detection with a smartphone. Lab on a Chip 12 (21) 4240-4243. https://doi.org/10.1039/c2lc40741h [ Links ]
SRIVASTAVA R, VERMA A and JAIN VK (2012) Low cost, disposable colorimetric sensor for quantitative detection of ammonia gas. Proc. SPIE 8549, 16th International Workshop on Physics of Semiconductor Devices, 85491A (15 October 2012). https://doi.org/10.1117/12.925176 [ Links ]
THOMPSON J and MORRISON G (1951) Determination of organic nitrogen. Control of variables in the use of Nessler's Reagent. Anal. Chem. 23 (8) 1153-1157. https://doi.org/10.1021/ac60056a029 [ Links ]
VALENTE IM, OLIVEIRA HM, VAZ CD, RAMOS RM, FONSECA AJM, CABRITA ARJ and RODRIGUES JA (2017) Determination of ammonia nitrogen in solid and liquid high-complex matrices using one-step gas-diffusion microextraction and fluorimetric detection. Talanta 167 747-753. https://doi.org/10.1016/j.talanta.2017.01.091 [ Links ]
VERDOUW H, VAN ECHTELD CJA and DEKKERS EMJ (1978) Ammonia determination based on indophenol formation with sodium salicylate. Water Res. 12 (6) 399-402. https://doi.org/10.1016/0043-1354(78)90107-0 [ Links ]
WIESMANN U (1994) Biological nitrogen removal from wastewater. In: Thomas S (ed.) Biotechnics/Wastewater. Springer, Berlin. 113-154. https://doi.org/10.1007/BFb0008736 [ Links ]
WYSZECKI G (1968) Technical notes: Recent agreements reached by the colorimetry committee of the commission internationale de l'eclairage. J Opt. Soc. Am. 58 (2) 290-292. https://doi.org/10.1364/JOSA.58.000290 [ Links ]
YAMADA K, HENARES TG, SUZUKI K and CITTERIO D (2015) Paper‐based inkjet‐printed microfluidic analytical devices. Angew Chem. Int. Ed. Engl. 54 (18) 5294-5310. https://doi.org/10.1002/anie.201411508 [ Links ]
YANG H and CHENG H (2007) Controlling nitrite level in drinking water by chlorination and chloramination. Sep. Purif. Technol. 56 (3) 392-396. https://doi.org/10.1016/j.seppur.2007.05.036 [ Links ]
Correspondence:
Phumlane Selby Mdluli
Email: phumlaneM@dut.ac.za
Received: 8 April 2019
Accepted: 2 June 2020